Why this happens The sensor is very sensitive and picks up every vibration on the chest, not only the heartbeat. When too many other vibrations are recorded, they mix with the heart signal and the recording becomes too noisy to analyse. Typical sources are talking, small movements, clothing rubbing against the sensor, and vibrations from the room or nearby equipment.
How to fix it
Move clothing, cables and anything else away from the sensor.
Press the patch firmly onto the skin. If it has lost its grip, use a fresh one.
Ask the patient, and everyone in the room, to stay silent during the recording.
Take a new measurement while the patient lies completely still.
Avoid it next time
Measure away from sources of vibration.
Don’t lean on or touch the surface the subject is lying on during the recording.
Still seeing the error? Try a different room to rule out vibrations from the building. If that doesn’t help, email support@ventriject.com and include:
Your sensor’s serial number (SN)
The date and time of the measurement
A screenshot of the error, if possible
Still need help?
Need additional assistance? Contact our support team at support@ventriject.com or submit a request through the self-service portal. For detailed operating guidance, please refer to the latest Seismofit® Sensor Instructions for Use (IFU).