Why this happens Seismofit® is designed for measurements taken lying down, so other positions do not give a reliable result. If the upper body is raised, the subject lies on their side, or they shift position during the recording, the app shows this error instead of a result.
How to fix it
Lay the subject flat on their back.
Lower any raised backrest so the upper body is level. If you use a pillow, make sure it is flat and keeps the head raised less than 45°.
Check the position before you start the recording.
Take a new measurement and ask the subject not to move until it is finished.
Avoid it next time
Use a firm, flat surface, such as an examination couch.
Start the recording only once the subject is settled and comfortable.
Still seeing the error? If the subject was lying flat and still the whole time, email support@ventriject.com and include:
Your sensor’s serial number (SN)
The date and time of the measurement
A screenshot of the error, if possible
Still need help?
Need additional assistance? Contact our support team at support@ventriject.com or submit a request through the self-service portal. For detailed operating guidance, please refer to the latest Seismofit® Sensor Instructions for Use (IFU).